GF303B three phase portable power source is a powerful and portable three phase current and voltage source All test values are generated absolutely synthetically with a high degree of accuracy and stability. It is high precision three phase energy meter calibration phantom load power source. GF303B uses the advanced synchronous sampling technology and electronic power amplifier components. The test equipment can generate sine-wave voltage/current and phantom power to drive energy meter testing. It is suitable for all the power engineering service company, electric power company distribution network automation departments, is also suitable for petrochemical, coal mine, railway and steel industry etc, to power automation departments use in the laboratory or on site.
Electrical parameters | |
---|---|
Accuracy class | 0.05%, 0.1% |
Power supply | Single phase AC 85V-265V, 50/60 Hz |
Communication Port | RS232, USB |
AC Voltage output | |
Range (U1, U2, U3 phase) | 0-300V |
Adjustment resolution | 0.01%, 0.1%, 1%, 10% |
Accuracy | 0.05% RG |
Stability | 0.01% RG / 1 min |
Distortion degree | Better than 0.1% (not capacitive load) |
Load Capacity | 15VA |
Full load regulation rate | 0.02% RG |
Full load regulation time | Less than 1mS |
Temperature drift | 16 PPM/°C |
Long-term stability | 60 PPM/year |
AC Current output | |
Range (I1, I2, I3 phase) | 0-6A |
Adjustment resolution | 0.01%, 0.1%, 1%, 10% |
Accuracy | 0.05% RG |
Stability | 0.01% RG/1min |
Distortion degree | Better than 0.1% (not capacitive load) |
Load capacity | 15VA |
Full load regulation rate | 0.02% RG |
Full load regulation time | Less than 1mS |
Temperature drift | 16 PPM/°C |
Long-term stability | 60 PPM/year |
Power output | |
Active power accuracy | 0.05% |
Reactive power accuracy | 0.1% |
Stability | 0.01% RG/1min |
Phase | |
Range | 0°-359.99° |
Adjustment resolution | 0.01°, 0.1°, 1°, 10° |
Accuracy | 0.05° |
Power factor | |
Adjusting range | -1 ~ 0 ~ + 1 |
Resolution | 0.0001 |
Accuracy | 0.0005 |
Frequency | |
Range | 40-65 Hz |
Adjustment resolution | 0.001 Hz, 0.01 Hz, 0.1 Hz, 1Hz |
Accuracy | 0.002 Hz |
Temperature drift | 0.5 PPM/°C |
Long-term stability | 4 PPM/year |
Harmonic content (compared with fundamental wave) | |
Voltage | ≤40% |
Current | ≤40% |
Phase | 0°-360° |
Harmonic | |
Harmonic times | 2-50 |
Adjustment resolution | 0.1% (compared with fundamental wave) |
Functions | |
Communication Port | RS232 |
Programmable controlled | Yes |
Harmonic | Yes |
Wiring mode | 3P4W, 3P3W, 1P2W |
Key | 40pcs |
LCD | 6 inch blue LCD |
PC control software | Optional |
Standard | |
Standard | IEC 62053-21,22, 23; IEC 60736; ANSI C12.20-2002; JJG 597-2005; JJG596-2012; JJG 1085-2013; JJF 68-2019; DL/T 826-2002; DL/T 1478-2015; DL/T 448-2016; JJG 51-1999; DL/T 2213.1-2021; GJB J 5857-2006; JJG6-2011; JJG70-2015; |
Safety | |
Isolation protection | IEC 61010-1:2001 |
Measurement Category | 300 V CAT III, 600 V CAT II |
Degree of protection | IP20 |
Declaration of conformity | CE & CNAS certified |
Mechanical parameters | |
Dimensions (W×D×H) (mm) | 300x400x120 |
Weight (kg) | 6.8 |
Environmental conditions | |
Working temperature | 0°C to 50°C |
Storage conditions | -30°C to 60°C |
Relative humidity | ≤85% |